Skip to main content

Full text of "USPTO Patents Application 10687850"

See other formats


(12) INTERNATIONAL APPLICATION PUBLISHED UNDER THE PATENT COOPERATION TREATY (PCT) 



(19) World Intellectual Property Organizatio 

International Bureau 



IIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIII 



(43) International Publication Date (10) International Publication Number 

6 March 2003 (06.03.2003) PCT WO 03/019166 Al 



(51) International Patent Classification 7 : 

// 33/49, A61B 5/053 



G01N 27/02 (74) Agent: DR LUDWIG BRANN PATENTBYRA AB; 

(Drottninggatan 7), Box 1344, S-75143 Uppsala (SB). 



(21) International Application Number: PCT/SE02/01530 

(22) International Filing Date: 27 August 2002 (27.08.2002) 

(25) Filing Language: English 

(26) Publication Language: English 

(30) Priority Data: 

0102896-8 29 August 2001 (29.08.2001) SE 

(71) Applicant (for all designated States except US): HAEMO 
WAVE AB [SE/SE]; Box 840, S-98228 Gallivare (SE). 

; (72) Inventors; and 

j (75) Inventors/Applicants (for US only): KASTEBO, Ove 

! [SE/SE]; Hedparken 17, S-98235 Gallivare (SE). KOIT- 

j SALU, Evald [SE/SE] ; SKansbergsvagen 27, S-14170 

j Huddingc (SE). NILSSON, Bertil [SE/SE]; Karabyvaacn 

! 285, S-24471 Dosjebro (SE). 



(81) Designated States (national): AE, AG, AL, AM, AT (util- 
ity model), AT, AU, AZ, BA, BB, BG, BR, BY, BZ, CA, 
CH, CN, CO, CR, CU, CZ (utility model), CZ, DE (util- 
ity model), DE, DK (utility model), DK, DM, DZ, EC, EE 
(utility model), EE, ES, FI (utility model), FI, GB, GD, GE, 
GH, GM, HR, HU, ID, IL, IN, IS, JP, KE, KG, KP, KR, KZ, 
LC, LK, LR, LS, LT, LU, LV, MA, MD, MG, MK, MN, 
MW, MX, MZ, NO, NZ, OM, PH, PL, PT, RO, RU, SD, 
SE, SG, SI, SK (utility model), SK, SL, TJ, TM, TN, TR, 
TT, TZ, UA, UG, US, UZ, VC, VN, YU, ZA, ZM, ZW. 

(84) Designated States (regional): ARIPO patent (GH, GM, 
KE, LS, MW, MZ, SD, SL, SZ, TZ, UG, ZM, ZW), 
Eurasian patent (AM. AZ. BY, KG, KZ, MD, RU, TJ, 'I'M), 
European patent (AT, BE, BG, CH, CY, CZ, DE, DK, EE, 
ES, FI, FR, GB, GR, IE, IT, LU, MC, NL, PT, SE, SK, 
TR), OAPI patent (BF, BJ, CF, CG, CI, CM, GA, GN, GQ, 
GW, ML, MR, NE, SN, TD, TG). 

Published: 

wiih international search report 



[Continued on next page] 



: (54) Title: SYSTEM AND METHt )D F( )R BLOOD ANALYSIS 




(57) Abstract: A method for blood analysis, comprising detecting the imaginary part of the complex impedance in a blood sam- 
pie having an unknown haemoglobin concentration, and directly correlating the imaginary part of the complex impedance with the 
haemoglobin concentration in said blood sample. 



WO 03/019166 Al llllllllllllllllllllllllllllllllllllllllllllllllll 



For two-letter codes and other abbreviations, refer to the "Guid- 
ance Notes on Codes and Abbreviations " a/ >pearing al the begin- 
ning of each regular issue of the PCT Gazette. 



WO 03/019166 



PCT/SE02/01530 



1 

SYSTEM AND METHOD FOR BLOOD ANALYSIS 

The Field of the Invention 

5 The present invention relates to a system and a method for measuring the 

haemoglobin value in blood, more specifically to measurements in a closed system. 

Background of the Invention 

10 Today's techniques for measuring the haemoglobin value in blood often lead to 

variations in the results of measurement since the analyses are performed on very 
small blood volumes (approx. 5-20 jul) , and therefore the operator's sampling 
technique is important for the results of the measurement. 

15 Furthermore, these known methods often involve a health risk for the operator, due 
to exposure to chemicals and contagious blood. 

The conventional methods also involve high costs for disposable material and for 
handling waste of disposable material and chemicals. 

20 

Hematocrit values obtained by conventional analyses can most of the time be 
correlated to the haemoglobin value, but they fail when measuring blood from 
patients suffering from certain blood anomalies which may cause the blood to 
contain abnormally large or small blood cells. 

25 

Methods for measuring hematocrit values are sometimes used when determining a 
state of anaemia and are much simpler than a method for measuring the oxygen- 
carrying component, haemoglobin, directly (see US-4 547 735, EP-0 417 796 and 
JP-03 099 254). 

30 

The above prior arts methods involve measuring the impedance in blood. 



35 



Traditional techniques for measuring the hematocrit value involve measuring the 
packed cell mass in a sample tube containing a blood sample. This measurement is 
performed after the blood sample having been centrifuged. 



WO 03/019166 



PCT/SE02/01530 



2 

In Ackmann et al, Specific impedance of canine blood, Ann. Biomed. Eng., 24 (1), 58 
(1996), measurements are disclosed where the imaginary part of the complex 
impedance is correlated with hematocrit value. Frequency intervals of 5 kHz to 1 
5 MHz have been used when analysing canine blood. This measuring technique 
involves non-direct measurement of the haemoglobin value. 

WO/ 0009996 or its equivalent US-5 792 668 discloses specific determination of 
glucose in NaCl using a radio frequency spectral analysis method. This method 
1 0 comprises analysing the real and the imaginary part of the complex impedance in 
the radio frequency range up to 5 GHz. 

Summary of the Invention 

15 Thus, there is a great need for a method for directly measuring the haemoglobin 

value in blood which does not pose a health risk to the user, does not involve high 
costs for handling of the disposable material used and enables measurements of 
small volumes of blood with accuracy. 

20 In view of the drawbacks associated with prior art methods it is therefore the object 
of the present invention to provide a system and a method that yields accurate 
haemoglobin results for small blood samples. This object is achieved with a method 
according to claim 1 and a system according to claim 9. 

25 The inventors have developed a technique primarily for determining the 

haemoglobin value in human blood in a closed system. Thus, the present invention 
permits safe measurements of the haemoglobin value in blood, eliminating the 
user's exposure to chemicals and contagious blood. 

30 The present invention also provides a system that presents high reliability in 

measurement results and that comprises a procedure of analysis that is so simple 
that only minimal laboratory experience is needed. 

Furthermore, the present invention rmnimises the production of environmentally 
35 detrimental waste like plastics and chemicals. 



WO 03/019166 



PCT/SE02/01530 



3 

The aim of the invention is also to provide a method that does not involve addition 
of agents that degrade the blood samples, so that an analysed blood sample can be 
used for further analyses at for example a central laboratory. 

5 

Brief description of the drawings 

The present invention will now become more fully understood from the detailed 
description given herein, wherein reference is made to the accompanying drawings, 
10 in which, 

Fig. 1 shows a scheme over the haemoglobin measuring system according to the 
present invention. 

1 5 Fig. 2 shows a calibration curve that illustrates the measured reference 

haemoglobin values as a function of its corresponding imaginary part of the 
complex impedance (values deriving from measurements performed at 400 kHz) . 

Fig. 3 shows a graph correlating the measured reference haemoglobin values and 
20 the haemoglobin values calculated using the inventive method (values originating 
from measurements of the imaginary part of the complex impedance performed at 
400 kHz). 

Fig. 4 shows a calibration curve, which shows the measured reference haemoglobin 
25 values as a function of its corresponding imaginaiy part of the complex impedance 
(values deriving from measurements performed at 500 kHz). 

Fig. 5 shows a graph correlating the measured reference haemoglobin values and 
the haemoglobin values calculated using the inventive method (values originating 
30 from measurements of the imaginary part of the complex impedance performed at 
500 kHz). 

Fig. 6 shows a calibration curve, which shows the measured reference haemoglobin 
values as a function of its corresponding imaginary part of the complex impedance 
35 (values deriving from measurements performed at 600 kHz). 



WO 03/019166 



PCT/SE02/01530 



4 

Fig. 7 shows a graph correlating the measured reference haemoglobin values and 
the haemoglobin values calculated using the inventive method (values originating 
from measurements of the imaginary part of the complex impedance performed at 
5 600 kHz). 

Fig. 8 shows a calibration curve, which shows the measured reference haemoglobin 
values as a function of its corresponding imaginary part of the complex impedance 
(values deriving from measurements performed at 700 kHz) . 

10 

Fig. 9 shows a graph correlating the measured reference haemoglobin values and 
the haemoglobin values calculated using the inventive method (values originating 
from measurements of the imaginary part of the complex impedance performed at 
700 kHz). 

15 

Fig. 10 shows a calibration curve, which shows the measured reference 
haemoglobin values as a function of its corresponding imaginary part of the 
complex impedance (values deriving from measurements performed at 800 kHz). 

20 Fig. 1 1 shows a graph correlating the measured reference haemoglobin values and 
the haemoglobin values calculated using the inventive method (values originating 
from measurements of the imaginary part of the complex impedance performed at 
800 kHz). 

25 Fig. 12 shows a calibration curve, which shows the measured reference 

haemoglobin values as a function of its corresponding imaginary part of the 
complex impedance (values deriving from measurements performed at 900 kHz) . 

Fig. 13 shows a graph correlating the measured reference haemoglobin values and 
30 the haemoglobin values calculated using the inventive method (values originating 
from measurements of the imaginary part of the complex impedance performed at 
900 kHz). 



WO 03/019166 



PCT/SE02/01530 



5 

Detailed Description of the Invention 

When directly measuring the haemoglobin value in the blood sample according to 
the present invention, one uses the fact that the blood has an alternating current 
5 (A.C.) impedance constituted by a resistive and a reactive part, where the reactive 
part is constituted by the imaginary part of the complex impedance. 

The above-mentioned prior art hematocrit measurements differ from the method 
according to the present invention in that they do not analyse the resistive and/ or 
10 the reactive part separately. 

If a molecule is exposed to an alternating electrical field, e.g. between two electrodes 
constituting a capacitor, it is influenced such that a capacity change occurs. 

15 The reactive part of the alternating current impedance of the blood is dependant on 
the amount of haemoglobin in the blood sample. The electrical complex impedance 
of the blood sample can then be measured and the magnitude of the reactive part, 
i.e. the imaginary part of the complex impedance, can be correlated with the 
haemoglobin value in the blood. 

20 

In the measurements according to the above-mentioned publication by Ackmann 
the imaginary part of the complex impedance is correlated with hematocrit value 
and not haemoglobin value, i.e. haemoglobin is not measured directly. Ackmann 
uses frequency intervals of 5 kHz to 1 MHz when analysing canine blood. 

25 

The present inventors have now identified frequency ranges where correlation 
between the haemoglobin value and the imaginary part of the complex impedance 
exists at least over a certain range in Hb concentration. 

30 The results described herein have been obtained by exciting haemoglobin molecules 
in blood samples by applying an alternating field within a range of 400-900 kHz on 
the blood sample using two electrodes. This frequency range partly overlaps the 
frequency range that Ackmann uses. 



WO 03/019166 



PCT/SE02/01530 



6 

There are thus several differences between this method and the method according 
to the present invention. 

The system according to the present invention will be described with reference to 
Fig. 1, in which it is generally described with 1. The system includes electrodes 204, 
5 an impedance meter 205, a display unit 210, a keyboard 211 and a memory 
(RAM/ROM) 212. 

The impedance meter 205 comprises a signal generator 206 and a signal-processing 
unit 207. The signal generator is capable of delivering an alternating current in a 
10 frequency range of 50 - 1200 kHz and measuring the impedance at one or more 
frequencies within the frequency range 50 - 1200. kHz. The signal-processing unit 
207 delivers an out-signal in analogue or digital form. 

In Fig. 1 there is also shown a sample tube 202 containing a blood sample 203. A 
15 septum, a rubber cork or the like seals the tube. 

To obtain the raw data from said blood sample, a network analyser (Rohde 8b 
Schwarz ZVC) including some additional equipment belonging to it and a measuring 
device was used. In the development work at The Technical University of Lulea. the 
20 vector network analyser that was used contained a signal generator 206 and a 
signal-processing unit 207 (see Fig. 1). 

The electrodes are preferably made of a material that does not oxidise, has good 
conductivity and does not affect the blood sample, e.g. platinum. 
25 Said electrodes are arranged in a tightly sealed tube, preferably in an ordinary 
sample tube. 

The system according to the invention uses only two electrodes 204, a measuring 
electrode and a reference electrode, but any number of electrodes can be employed 
30 in combination. 

To prevent air bubbles from interfering with the measuring process, the electrodes 
204 of the measuring device are introduced from below into the sealed blood sample 
tube 202 mounted upside-down, so that the electrodes penetrate the sealing of the 



WO 03/019166 



PCT/SE02/01530 



7 

blood sample tube and come into direct contact with the blood without first passing 
through air. 

The signal generator 206 generates the electrical alternating field, which is applied 
5 to the blood sample 203 via the electrodes 204, the electrical alternating field 
preferably having a frequency within the range 50 - 1200 kHz. 

The signal from the electrodes is amplified and filtered by the signal-processing unit 
207, as a preparation for the transformation from analogue to digital form. 

10 

After filtering and amplification the signal-processing unit 207 processes the signal 
mathematically, whereby the reactive part of the signal is correlated with the 
haemoglobin value of the blood. This is performed within the frequency range 50 - 
1200 kHz, and preferably at about 800 kHz, since the best correlation has been 
15 obtained at this frequency. 

This information is subsequently prepared for presentation on the display unit 210. 
From the keyboard 211a user can interact with the system, such as feeding it with 
patient information and data, in addition to controlling the system. 

20 

Furthermore, the signal-processing unit 207 is coupled to a memory 212, which 
preferably comprises a RAM and a ROM, wherein measurement data and other 
information can be saved and read. 

25 The method according to the present invention is based on the discovery that within 
certain frequency intervals correlation exists between the haemoglobin value in 
blood within a certain concentration range and its imaginary part of the complex 
impedance. 

30 In order to perform the method according to the invention, a standard curve must 
be constructed. The imaginary part of the complex impedance is then measured 
and correlated with the haemoglobin value using the standard curve. 

The standard curve is constructed by first centrifuging a reference blood (arbitrary 
35 blood sample) to obtain plasma. The plasma is then used to dilute the reference 



WO 03/019166 



PCT/SE02/01530 



8 

blood to five appropriate concentrations in haemoglobin. These concentrations lie in 
the range of normal human haemoglobin values. This interval is chosen because it 
matches the interval in which the ordinary measurement will be performed, but it is 
also chosen because linearity exists in this interval, at least for haemoglobin values 
5 of 80 to 180. 

The haemoglobin values in these samples are then measured with a reference 
instrument. 

10 The respective corresponding imaginary parts of the complex impedance are 

determined at a frequency within the frequency range 50 - 1200 kHz using the 
above-mentioned network analyser (Rohde & Schwarz ZVC) . 

The impedance results, obtained at the chosen frequency, are correlated with the 
1 5 haemoglobin results obtained with the reference instrument. The equation of a 

calibration curve within the above-mentioned range is determined. A ready-to-use 
system comprises a computer system having one or more of these equations stored 
in its memory unit. When performing the haemoglobin measurements the computer 
will then perform the correlation procedure. 

20 

The measurement of the imaginary part of the complex impedance in patient blood 
samples is performed by analysing blood samples from patients using the above- 
described measuring equipment (Rohde & Schwarz ZVC) at the same frequency 
within the frequency range 50 - 1200 kHz as mentioned above. 

25 

The results obtained from measuring the imaginary part of the complex impedance 
at the chosen frequency are fed into the equations of the respective calibration 
curve, yielding calculated corresponding haemoglobin values. 

30 The reproducibility of the described method was determined by analysing several 
times each of a number of patient samples with the measuring equipment. The 
variation expressed in %CV was 2,5 or lower at the level 80 - 135 g haemoglobin/1. 



WO 03/019166 PCT/SE02/01530 

9 

Examples 

The method will now be described with reference to non-limiting examples. 
5 Example 1 . Correlation studies of haemoglobin at 400 kHz 
Construction of a standard curve 

A reference blood sample (normal sample) was centrifuged and the plasma was 
10 separated in order to be used for dilution of the blood sample to 5 different levels of 
haemoglobin. 

The diluted samples were analysed with a reference instrument (SYSMEX SF-3000) 
at the levels 65, 85, 141, 151 and 179 g/1 considering the haemoglobin values. 

15 

An alternative way would be to provide a reference blood sample exhibiting a known 
haemoglobin concentration and then dilute said reference blood sample to obtain a 
set of reference samples having different known haemoglobin concentrations. 

20 Still another way would be to provide a plurality of reference blood samples 
exhibiting unknown haemoglobin concentrations and then determine the 
haemoglobin concentrations in said reference blood samples to obtain a set of 
reference samples having different known haemoglobin concentrations. 

25 The corresponding imaginary part of the complex impedance for each sample was 
determined at 400 kHz using the above-mentioned network analyser (Rohde & 
Schwarz ZVC). 

The impedance results obtained at 400 kHz were correlated with the haemoglobin 
30 results obtained with the reference instrument. The equation of a calibration curve 
within the mentioned range (65 - 179 g/1) was determined (see Fig. 2). 

The standard curve shows the haemoglobin values (Hb) of the above-mentioned 
blood samples, as obtained with the reference instrument, as a function of their 
35 corresponding imaginary parts of the complex impedance. 



WO 03/019166 



PCT/SE02/01530 



10 

The obtained values were correlated and the following equation was drawn up: 
Y = A + B*X 

5 where Y represents the haemoglobin values in the reference blood sample and X 
their corresponding imaginary parts of the complex impedance. 

The factor A and the coefficient B were found to be 32,62 and -1632, respectively. 

10 The R and the R-square values were 0.9955 and 0,9910, respectively (see Fig. 2). 

Measurement of a patient sample 

Subsequent to the construction of the standard curve 93 patient blood samples 
15 were analysed using both the reference instrument and the above-described 
measuring equipment at 400 kHz. 

The results obtained from measuring the imaginary part of the complex impedance 
were fed into the previously made calibration curve to transform them to 
20 haemoglobin values. 

The calculated haemoglobin values (X- values), ranging between 80-180 were 
correlated with the haemoglobin values from the reference instrument (Y- values) . 

25 It was found that the factor A and the coefficient B were 12,33 and 0,7868, 
respectively. 

The R and the R-square values were 0.8563 and 0,7333, respectively (see Fig. 3). 

30 Example 2. Correlation studies of haemoglobin at 500 kHz 

The standard curve for the measurement performed at 500 kHz was constructed in 
the same manner as in Example 1, which for this frequency resulted in A and B 
being 31,60 and -1451, respectively, and R and R-square being 0,9966 and 0,9932, 
35 respectively (see Fig. 4). 



WO 03/019166 



PCT/SE02/01530 



11 

The correlation studies, which also were performed in the same manner as in 
Example 1, resulted in A and B being 6,065 and 0,8441, respectively, and R and R- 
square being 0,8943 and 0,7998, respectively (see Fig. 5). 

5 

Example 3. Correlation studies of haemoglobin at 600 kHz 

The standard curve for the measurement performed at 600 kHz was constructed in 
the same manner as in Example 1, which for this frequency resulted in A and B 
10 being 30,67 and -1336, respectively, and R and R-square being 0,9974 and 0,9949, 
respectively (see Fig. 6). 

The correlation studies, which also were performed in the same manner as in 
Example 1, resulted in A and B being 6,119 and 0,8578, respectively, and R and R- 
15 square being 0,9030 and 0,8155, respectively (see Fig. 7). 

Example 4. Correlation studies of haemoglobin at 700 kHz 

The standard curve for the measurement performed at 700 kHz was constructed in 
20 the same manner as in Example 1, which for this frequency resulted in A and B 

being 29,88 and -1255, respectively, and R and R-square being 0,9981 and 0,9961, 
respectively (see Fig. 8). 

The correlation studies, which also were performed in the same manner as in 
25 Example 1, resulted in A and B being -0,7240 and 0,9136, respectively, and R and 
R-square being 0,9453 and 0,8937, respectively (see Fig. 9). 

Example 5. Correlation studies of haemoglobin at 800 kHz 

30 The standard curve for the measurement performed at 800 kHz was constructed in 
the same manner as in Example 1, which for this frequency resulted in A and B 
being 29,42 and -1200, respectively, and R and R-square being 0,9984 and 0,9968, 
respectively (see Fig. 10). 



WO 03/019166 



PCT/SE02/01530 



12 

The correlation studies, which also were performed in the same manner as in 
Example 1, resulted in A and B being -1,120 and 0,9088, respectively, and R and R- 
square being 0,9478 and 0,8983, respectively (see Fig. 11). 

5 Example 6. Correlation studies of haemoglobin at 900 kHz 

The standard curve for the measurement performed at 900 kHz was constructed in 
the same manner as in Example 1 , which for this frequency resulted in A and B 
being 29,13 and -1 155, respectively, and R and R-square being 0,9986 and 0,9971, 
10 respectively (see Fig. 12). 

The correlation studies, which also were performed in the same manner as in 
Example 1, resulted in A and B being 3,189 and 0,9021, respectively, and R and R- 
square being 0,9470 and 0,8967, respectively (see Fig. 13). 

15 

It should be understood that the detailed description and specific examples, while 
indicating preferred embodiments of the invention are given by way of example only. 
Various changes and modifications within the spirit and scope of the invention will 
become apparent to those skilled in the art from this detailed description. 



WO 03/019166 



PCT/SE02/01530 



A method for blood analysis, comprising 

detecting the imaginary part of the complex impedance in a blood sample 
having an unknown haemoglobin concentration, and 

directly correlating the imaginary part of the complex impedance with the 
haemoglobin concentration in said blood sample. 

Method for blood analysis according to claim 1 , wherein the imaginary part 
of the complex impedance is measured within the frequency range 50-1200 
kHz, preferably at about 800 kHz. 

Method for blood analysis according to claim 2, wherein the imaginary part 
of the complex impedance is measured at one or more frequencies within 
said frequency range. 

Method for blood analysis according to any of claims 1-3, which prior to the 
step of detecting the imaginary part of the complex impedance, comprises the 
further steps of: 

providing a reference blood sample exhibiting an unknown haemoglobin 
concentration; 

diluting said blood reference sample to obtain a set of reference blood 
samples having unknown haemoglobin concentrations; 

determining the haemoglobin concentrations in said reference blood samples 
to obtain a set of reference blood samples having known haemoglobin 
concentrations ; 



35 



applying an alternating current (A.C.) electrical field to each reference blood 
sample using electrodes in direct contact with the blood, 



WO 03/019166 



PCT/SE02/01530 



14 

measuring the imaginary part of the complex impedance of each of said 
reference blood sample, at one or more frequencies within the frequency 
range 50 - 1200 kHz, 

correlating the imaginary part of the complex impedance with the 
haemoglobin value of the blood in each of said reference samples to obtain 
one or more standard curves. 

Method for blood analysis according to any of claims 1-3, which prior to the step of 
detecting the imaginary part of the complex impedance, comprises the further steps of: 

providing a plurality of reference blood samples exhibiting unknown 
haemoglobin concentrations; 

determining the haemoglobin concentrations in said reference blood samples 
to obtain a set of reference samples having different known haemoglobin 
concentrations; 

applying an alternating current (A.C.) electrical field to each reference blood 
sample using electrodes in direct contact with the blood, 

measuring the imaginary part of the complex impedance of each of said 
reference blood sample, at one or more frequencies within the frequency 
range 50 - 1200 kHz, 

correlating the imaginary part of the complex impedance with the 
haemoglobin value of the blood in each of said reference samples to obtain 
one or more standard curves. 

Method for blood analysis according to any of claims 1-3, which prior to the 
step of detecting the imaginary part of the complex impedance, comprises the 
further steps of: 



35 



providing a reference blood sample exhibiting a known haemoglobin 
concentration; 



WO 03/019166 PCT/SE02/01530 

15 

diluting said reference blood sample to obtain a set of reference samples 
having different known haemoglobin concentrations; 

5 applying an alternating current (A.C.) electrical field to each reference blood 

sample using electrodes in direct contact with the blood, 

measuring the imaginary part of the complex impedance of each of said 
reference blood sample, at one or more frequencies within the frequency 
10 range 50 - 1200 kHz, 

correlating the imaginary part of the complex impedance with the 
haemoglobin value of the blood in each of said reference samples to obtain 
one or more standard curves. 

15 

7. Method for blood analysis according to any of claims 4-6, whereby the 
correlation comprises feeding the result obtained from measuring the 
imaginary part of the complex impedance of a blood sample having an 
unknown haemoglobin concentration at the chosen frequency/ frequencies 

20 into said calibration curve /curves, to obtain its corresponding calculated 

haemoglobin concentration based on the imaginary part of the complex 
impedance of said blood sample. 

8. Method for blood analysis according to claim 1, comprising 

25 

measuring the impedance in a blood sample having an unknown 
haemoglobin concentration using an impedance meter at one or more 
frequencies within the frequency range 50 - 1200 kHz. 

30 9. A system for blood analysis, comprising 



at least two electrodes (204) adapted to be introduced into a blood sample 
(203); 



WO 03/019166 



PCT/SE02/01530 



16 

an impedance meter (205) coupled to said electrodes (204) and capable of 
delivering an alternating current in a frequency range of 50 - 1200 kHz, so as 
to enable measurement of impedance at one or more frequencies within said 
frequency range, and to deliver an out-signal in analogue or digital form; 

a signal processing unit (207) for processing the impedance signal and 
calculating a haemoglobin value on the basis of said measured impedance. 



o O o 



WO 03/019166 



PCT/SE02/01530 



1/7 




WO 03/019166 



PCT/SE02/01530 




WO 03/019166 



PCT/SE02/01530 




WO 03/019166 



PCT/SE02/01530 



4/7 



■ • Data -t-Y = 30,6693- 1335, 53'Xj 

;— 95% Confidence (Data) — «— 95% Confidence (Line) I 









2CO-- 


• 'it.;.-:.- : 








180- 












160- 

140- 

120- 


































100- 

80 












60- 












40- 
























20- 



-0,12 -0,1 -0,08 -0,06 -0,04 -0,02 0 

Im j 

i 

Fig. 6 



| * Data --7— Y = 6,1 1897+ 0,85777*X 

j — «— 95% Confidence (Data) —a— 95% Confidence (Line) 

























"4 - ^ 











80 100 120 140 160 180 200 



Fig. 7 



WO 03/019166 



PCT/SE02/01530 




Fig. 9 



WO 03/019166 



PCT/SE02/01530 



6/7 



• Data — ▼ — Y = 29,41 53- 1 200,29*X 
95% Confidence (Data) — «— 95% Confidence (Line) 



5^ . 








24 0- 

-190- 

-1-70-- 


• 






-150- 

-130- 












-140- 

90- 










70- 

50- 



-0,14 -0,12 -0,1 -0,08 -0,06 -0,04 -0,02 



Fig. 10 



j • Data -r—Y= 1,11953 + O,90875*X 

j — ♦- 95% Confidence (Data) - a - 95% Confidence (Line) 

200 i : 




80 100 120 140 160 180 200 

Calc. Hb 



Fig. 11 



WO 03/019166 



PCT/SE02/01530 




Fig. 13 



INTERNATIONAL SEARCH REPORT 



I Interna 

I PCT/i 



I Internatcoiial application No. 

PCT/SE 02/01530 



A. CLASSIFICATION OF SUBJECT MATTER 



IPC7: G01N 27/02 // G01N 33/49, A61B 5/053 

According to International Patent Classification (IPC) or to both national classification and IPC 



B. FIELDS SEARCHED 



Minimum documentation searched (classification system followed by classification symbols) 

IPC7: G01N 



Documentation searched other than minimum documentation to the extent that such documents are included in the fields searched 

SE,DK,FI,N0 classes as above 



Electronic data base consulted during the international search (name of data base and, where practicable, search terms used) 

WPI DATA, EPQ INTERNAL, PAJ, MEDLINE, CAPLUS, INSPEC, BIOSIS, EMBASE, SCISE 
ARCH 



C. DOCUMENTS CONSIDERED TO BE RELEVANT 



Citation of document, with indication, where appropriate, of the relevant passages 



Relevant to claim No. 



EP 0417796 A2 (KABUSHIKI KAISHA TOYOTA CHUO 
KENKYUSHO) , 20 March 1991 (20.03.91), 
abstract 



Med. Biol. Eng. Comput. vol. 37, 1999, 

M.Y. Jaffrin et al : "Comparison of optical, 
electrical, and centr if ligation techniques 
for haematocrit monitoring of dlalysed 
patients", pages 433-439, abstract 



| | Further documents are listed in the continuation of Box C. | yj See patent family a 



Special categories of cited 

it defining the general 



te of ft 



be of particular relevance 
"E" earlier application or patent but published on or after the international 
filing date 

"L" document which may throw doubts on priority claims) or which is 
cited to establish the publication date of another citation or other 
special reason (as specified) 

"O" document referring to an oral disclosure, use, exhibition or other 

al filing date but later than 



"T" later document published after the international filing date or priority 
date and not in conflict with the application but cited to understand 
the principle or theory underlying the invention 

"X" document of particular relevance: the claimed invention cannot be 
considered novel or cannot be considered to involve an inventive 
step when the document is taken alone 

" Y" document of particular relevance: the claimed invention cannot be 
considered to involve an inventive step whe- - ' - 



being obvious to a person skilled in the art 
ber of the same patent family 



Date of the actual completion of the international search 



3 December 2002 



Date of mailing of the international search report 

0 4-12- 2002 



Name and mailing address of the ISA/ 
Swedish Patent Office 
Box 5055, S-102 42 STOCKHOLM 
Facsimile No. + 46 8 666 02 86 



Authorized officer 



Jens Waltin/Els 

Telephone No. +46 8 782 25 00 



Form PCT/ISA/210 (second sheet) (July 1998) 



INTERNATIONAL SEARCH REPORT 

Information on patent family n 



InternasTBnal application No. 

PCT/SE 02/01530 



0417796 A2 20/03/91 



SE 0417796 T3 

DE 69014262 D,T 20/07/95 

JP 2665806 B 22/10/97 

JP 3099254 A 24/04/91 



Form PCT/ISA/210 (patent family annex) (July 1998) 



DERWENT-ACC-NO: 2003-300754 



DERWENT— ACC— NO : 2003-300754 
DERWENT— WEEK : 20 0452 

COPYRIGHT 2008 DERWENT INFORMATION LTD 

TITLE: Blood analysis for measuring hemoglobin 

value, involves detecting imaginary part 
of complex impedance in blood sample 
having unknown hemoglobin concentration, 
and directly correlating imaginary part 
with hemoglobin concentration 

INVENTOR: KASTEBO, 0; KOITSALU, E ; NILSSON, B 

PATENT-ASSIGNEE: HAEMO WAVE AB [HAEMN] 

PRIORITY-DATA: 2001SE-0002896 (August 29, 2001) 

PATENT-FAMILY : 

PUB-NO PUB— DATE LANGUAGE PAGES MAIN-IPC 

AU 2002327988 Al March 10, N/A 000 G01N 027/02 

2003 

WO 2003019166 Al March 6, E 027 G01N 027/02 
2003 

SE 200102896 A March 1, N/A 000 G01N 027/02 
2003 

SE 521208 C2 October 14, N/A 000 G01N 027/02 
2003 



file:///CI/Documenl , 20and' J Settings/ ASoderquistAIy... l_2008-02-02_WO_2003019166_A_M_AccessibleVersion.htm (1 of 4)2/2/2008 3:39:39 PM 



DERWENT-ACC-NO: 2003-300754 



DESIGNATED— STATES : 



AE AG AL AM AT AU AZ BA BB BG BR BY BZ 

CA CH CN CO CR CU CZ DE DK DM DZ EC EE 

ES FI GB GD GE GH GM HR HU ID IL IN IS 

JP KE KG KP KR KZ LC LK LR LS LT LU LV 

MA MD MG MK MN MW MX MZ NO NZ OM PH PL 

PT RO RU SD SE S G SI SK SL TJ TM TN TR 

TT TZ UA UG US UZ VC VN YU ZA ZM ZW AT 

BE BG CH CY CZ DE DK EA EE ES FI FR GB 

GH GM GR IE IT KE LS LU MC MW MZ NL OA 

PT SD SE SK SL SZ TR TZ UG ZM ZW 



APPLICATION-DATA : 



APPL-DESCRIPTOR APPL-NO 



AU2002327988A1 N/A 



AU2002327988A1 Based on 
WO2003019166A1 N/A 



SE 200102896A N/A 



SE 521208C2 



N/A 



APPL- 
DATE 



2002AU-0327988 August 

27, 2002 
WO2003019166 N/A 
2002WO-SE01530 August 

27, 2002 
2001SE-0002896 August 

29, 2001 
2001SE-0002896 August 

29, 2001 



INT-CL (IPC) : A61B005/053, G01N027/02 , G01N033/49 

ABSTRACTED-PUB-NO: WO 200301916 6A 

BASIC-ABSTRACT: 

NOVELTY - Analysis of blood, includes: 

(1) detecting the imaginary part of complex impedance in a 
blood sample having unknown hemoglobin concentration; and 

(2) directly correlating the imaginary part of complex 
impedance with the hemoglobin concentration. 



DETAILED DESCRIPTION - An INDEPENDENT CLAIM is also 



file:///CI/Documenl , 2' > md' 2( > Settings/ASoderquist/My... l_2008-02-02_WO_2003019166_A_M_AccessibleVersion.htm (2 of 4)2/2/2008 3:39:39 PM 



DERWENT-ACC-NO: 2003-300754 



included for a blood analysis system comprising: 

(1) at least two electrodes (204) for introduction into a 
blood sample (203) ; 

(2) an impedance meter (205) coupled to the electrodes and 
capable of delivering an alternating current at 50 - 1200 
kHz to allow measurement of impedance at more than one 
freguencies within the range and to deliver an out-signal 
in analog or digital form; and 

(3) a signal processing unit (207) for processing the 
impedance signal and calculating hemoglobin value based on 
the measured impedance . 

USE - Analysis is used for measuring the hemoglobin value 
in blood sample. 

ADVANTAGE - The analysis permits safe measurements of 
hemoglobin value in blood, eliminating the user's exposure 
to chemicals and contagious blood. It does not involve high 
costs for handling of disposable material used, and enables 
measurement of small blood volume with accuracy. 

DESCRIPTION OF DRAWING (S) - The figure is a schematic view 
of a blood analysis system. 

Blood sample 203 

Electrodes 204 

Impedance meter 205 

Signal processing unit 207 

CHOSEN— DRAWING : Dwg .1/13 

TITLE-TERMS: BLOOD ANALYSE MEASURE HAEMOGLOBIN VALUE 

DETECT IMAGINARY PART COMPLEX IMPEDANCE 
BLOOD SAMPLE UNKNOWN HAEMOGLOBIN 
CONCENTRATE CORRELATE IMAGINARY PART 
HAEMOGLOBIN CONCENTRATE 



file:///CI/Documenl , 2' > md' 2( > Settings/ASoderquist/My... l_2008-02-02_WO_2003019166_A_M_AccessibleVersion.htm (3 of 4)2/2/2008 3:39:39 PM 



DERWENT-ACC-NO: 2003-300754 



DERWENT— CLASS : B04 P31 S03 



CPI-CODES: 



EPI-CODES: 
CHEMICAL-CODES : 



B04-B04D2; B04-B04D5; B11-C08D; B11-C09; 
B12-K04A; 

S03-E02; S03-E14H; S03-E14H1; 
Chemical Indexing Ml *01* 
Code M417 M423 M750 M905 
Compounds A04LJK A04LJA 



Fragmentation 
N102 Specfic 



Chemical Indexing M6 *02* Fragmentation 
Code M905 R501 R515 R528 R611 R637 



SECONDARY-ACC-NO : 

CPI Secondary Accession Numbers: C2003-078443 
Non-CPI Secondary Accession Numbers : N2003-239242 



file:///CI/Documenl , 20and' 20Settings/ASoderquist/My... l_2008-02-02_WO_2003019166_A_M_AccessibleVersion.htm (4 of 4)2/2/2008 3:39:39 PM